The first operational prototype of a high-purity germanium (HPGe) detector developed within the European LEAPS-INNOV project is presented in this work. This prototype features a monolithic, multi-element sensor optimized for high-resolution X-ray spectroscopy in the hard X-ray regime, intended for handing high count rates of 30 Mcps, across a broad energy range (5–100 keV). The present prototype demonstrates a throughput of about 150–200 kcps/pixel, corresponding to a total count rate of about 1.5–2 Mcps for the full 10 element detector. We discuss here a complete laboratory-based characterization of the detector's performance, as well as an on-beam evaluation at the BM05 beamline of the ESRF synchrotron facility, using monochromatic X-rays in the 20–50 keV energy range. We provide a detailed performance assessment that also includes a phenomenological defect-depth estimation model.
Development and first performance evaluation of multi-element monolithic HPGe detector for X-ray spectroscopy
Porro, M.;
2026
Abstract
The first operational prototype of a high-purity germanium (HPGe) detector developed within the European LEAPS-INNOV project is presented in this work. This prototype features a monolithic, multi-element sensor optimized for high-resolution X-ray spectroscopy in the hard X-ray regime, intended for handing high count rates of 30 Mcps, across a broad energy range (5–100 keV). The present prototype demonstrates a throughput of about 150–200 kcps/pixel, corresponding to a total count rate of about 1.5–2 Mcps for the full 10 element detector. We discuss here a complete laboratory-based characterization of the detector's performance, as well as an on-beam evaluation at the BM05 beamline of the ESRF synchrotron facility, using monochromatic X-rays in the 20–50 keV energy range. We provide a detailed performance assessment that also includes a phenomenological defect-depth estimation model.I documenti in ARCA sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.