The DePFET Sensor with Signal Compression (DSSC) instrument is a 1 Mpixel camera under development at the European XFEL for the detection of low-energy X-rays (<;6 keV). In this work an experimental study of charge collection effects has been carried out as a function of the deposited charge in order to extract the true time profile of the collected charge cloud in a wide range of experimental conditions. A preliminary set of results is presented and discussed.

Study of charge collection effects in the DSSC sensor

Porro M.;
2018-01-01

Abstract

The DePFET Sensor with Signal Compression (DSSC) instrument is a 1 Mpixel camera under development at the European XFEL for the detection of low-energy X-rays (<;6 keV). In this work an experimental study of charge collection effects has been carried out as a function of the deposited charge in order to extract the true time profile of the collected charge cloud in a wide range of experimental conditions. A preliminary set of results is presented and discussed.
2018
2017 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in ARCA sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10278/5008722